The MPS is a multifunctional autosampler and sample preparation robot for GC and GC/MS. The MPS provides modern analytical laboratories high throughput, high performance and unparalleled flexibility.
The following techniques are easily automated using the MPS:
Solid Phase Extraction SPE
- Disposable Pipette Extraction DPX
- Dynamic HeadSpace DHS
- Headspace HS
- Solid Phase Micro Extraction SPME
- Thermal desorption of solids or liquids in µ-vial inserts (ATEX)
- Stir Bar Sorptive Extraction SBSE (GERSTEL Twister®)
- Automated GC inlet Liner EXchange (ALEX)
- Liquid sample preparation – Sample Prep by mouse-click
- Heated or cooled sample conditioning and agitation
- Parallel evaporation and concentration (mVAP)
- Reading and processing of barcodes
- Integrated weighing option
https://www.youtube.com/watch?v=yZH5ntR5BYU
The MPS is easily and conveniently controlled from the GERSTEL MAESTRO Software – stand-alone or integrated with the Agilent Technologies ChemStation:
- Just One method and one sequence table controls the entire system
- Sample preparation is performed during GC analysis of the preceding sample ensuring highest possible throughput and system utilization
- Sample Prep by mouse-click
Multi-Purpose-Sampler-MPS Brochure
Product Specifications
| Sampler Type |
- XYZ robot with syringe-only concept
- positioning accuracy ± 0.1 mm
|
| System Configuration |
- compatible with most standard GCs
- compatible with all standard inlets
- operates with all injector position
- for more than one injector
- numerous accessories available
|
| System Version |
- MPS Liquid
- MPS Headspace
- MPS SPME
|
| Control |
- based on GERSTEL MAESTRO software and the MPS handheld keypad
- sample prep by mouse click
- in-depth, context sensitive online help for easy operation of the software
|
| Sample capacity |
- 200 x 1 mL vials per tray
- 98 x 2 mL vials per tray
- 78 x 1 mL vials per tray
- 32 x 10 mL/20 mL vials per tray
- 8 x 100 mL vials per tray
- max. 4 trays, depending on configuration
|
| Thermostated Sample Trays |
- Temperature range 4 … 200°C
- for 1 mL, 2 mL, 10 mL and 20 mL vials
- with peltier or liquid cooling
|
| Wash stations |
- solvents 2 … 4
- solvent reservoir 10 … 1000 mL
|
| Dilutor Module |
- 1 mL, 2.5 mL, 5 mL and 10 mL dilutor syringe
- solvent reservoir 1000 mL
|
| Fiber bake-out station |
- bake out temperature max. 350°C
- gas flow 6 or 12 mL/min
|
| Interfaces |
- 2 x AUX to connect heaters or agitators
- 2 x Interface
- RS 232 to connect a computer
- LAN to connect a computer
- WASH STATION
- MODBUS for valve control
|
| Operating Conditions |
- 4 … 40°C
- relative humidity 75%, non-condensing
|
| Power Requirements |
- 100 … 240 VAC
- 50/60 Hz
- 120 W
|
| Dimensions (H x W x D) |
64.8 x 82.8 x 38.5 cm |
| Weight |
10 kg (without accessories) |
| Extension & Options |
- GERSTEL SPE, for automated Solid Phase Extraction
- Twister-Option, for automated desorption of GERSTEL Twister inside the Thermal Desorption Unit
- DHS, for automated Dynamic Headspace analysis
- ATEX-Option, for Automated-liquid injection, thermal extraction of the liquid and tube exchange in a GERSTEL Thermal Desorption Unit (TDU)
- ALEX-Option, automated liner exchange for the GERSTEL CIS used for samples with a high matrix-load
- SPME Multi Fiber Exchange MFX, for the Solid Phase Micro Extraction with automated fiber exchange
- Disposable Pipette Extraction (DPX), a novel alternative to SPE with a powdery adsorbent
- MPS PrepStation, the system with 2 towers and 2 rails allows the simultaneous use of two different syringes or the combination of liquid sample prep and SPME or HS sample introduction
- special solutions on requst
|
MPS Liquid
| Vials |
|
| Syringe |
- 1.2, 5, 10 , 25, 100, 250, 500 or 1000 µL
- particularly LVI syringes allow slower injections
- heated 10 µL syringe, 35 … 80°C
- one universal syringe holder for all syringe volumes
|
| Injection Volume |
- 0.12 … 1000 µL
- max. 99 injectors per vial
|
| Injection Modes |
- standard
- sandwich technique using air solvent
- large volume
|
| Speed |
- injection speed 0.05 … 500 µL/s, depending on syringe volume
- fill speed 0.05 … 500 µL/s, depending on syringe volume
- separate settings for injection and rinse steps
- separate settings for sample and solvent
|
| Precision |
<0.8% RSD |
| Carryover |
<0.08% |
| Rinsing Steps |
- separate adjustment of pre- and post-clean steps
- using sample or solvent
- for up to 2 different solutions
|
| Needle Penetration |
- in sample vial 1 … 45 mm
- in solvent vial 1 … 45 mm
- in CIS injector 20 … 45 mm
|
| Specials with MAESTRO Software only |
- real sandwich technique
- standard edition
- multiple sample prep-ahead
|